Advanced Process Characterization for Lithography |
|
![]() |
|
|
The Inspector Product SeriesThe Inspector series of instruments are designed for real-time, in-line measurement of film thickness, reflectivity or transmissivity. Probes and probe mounts are available for many commercial coat/develop systems making in-situ measurement easier than ever before. Additional software application modules are available for determination of Develop Rate data. Optional Hardware includes computer controlled motorized scanning stages to make uniformity measurements quick and easy. A combined deuterium/halogen light source extends the usable wavelength range down to 190 nm for advanced DUV process development and characterization (Standard halogen illumination covers the 480 to 900 nm range). In-situ or off-line, thickness, reflection or transmission, UV-VIS, whatever you require, we can provide the perfect solution. |
|
|
ProLight Software with the LithoPak Applications Module |
|
|
Thickness
& Uniformity
Photoresist uniformity across an 8” wafer. Variation less than 2 Å are easily observed.. The intensity vs. wavelength provide both the measured film reflectivity as well as the best fit to the data. |
![]() |
| Trench
Depth
The DepthPak applications module determines the depth of trenches in silicon wafers. The technique is noninvasive and determines both the depth of the trench and the thickness of the top layer. The wafer to wafer repeatability of the shallow trench depth is improved by endpointing the trench etch with the Inspector depth measurement or monitoring stability of the process with the off-line trench depth measurements. |
![]() |
| Photoresist
Characterization
Recording time dependent data has never been easier. Determine the best bake time by watching the solvents evaporate. Measure develop time-to-clear to aid recipe development. Look at thickness vs time during develop to see if standing waves are present. Collect Thickness vs time plots of multiple exposures to input to our develop rate data analysis package, (DevPak). |
||
![]() |
![]() |
![]() |
| Clear-point is easily determined in this multi-wavelength Intensity vs Time plot. | These thickness vs time plots illustrate the effect of exposure dose on develop rate. | Thickness changes during a photoresist bake due to solvent evaporation. |
|
Accuracy
|
![]() Database Automatic data logging into the SQL compatible database make SCP data collection easier than ever before. Data can be copied to the clipboard or saved as .CSV files for easy analysis and charting. With an optional network card, the software can log the data to remote computers as well. |
| Triggering
& I/O
The Inspector can be triggered in three ways: Manually, electronically and optically. All digital I/O channels are fully optically isolated. The inputs are compatible with both TTL and contact closure, while the outputs provide fully isolated TTL for signaling process completion as well as warnings due to process variation. Recipes can even be selected with fully isolated digital inputs. The optical trigger detects the presence of a wafer by looking at intensity changes. Trigger delays can be set so that acquisition doesn’t start until the wafer has been lowered onto the chuck or stage and the process step starts. The combination of remote data logging with the optical trigger means that every wafer can be measured with no user input. |
|
![]() |
Intensity
Equations Mouse functions and pop-up menus make the creation of wavelength functions simple and easy. The display to the right shows the intensities of four wavelengths and the film thickness as a function of time. Normalization and offset were all done within the equation editor. Equations can be named, saved, and associated with specific channels of any recipe. |
| Hardware Features | |
| Illumination:
Standard illumination
is provided by a 10,000-hr lifetime halogen bulb, spanning a useful spectral
range from 480 to 900 nm. A 480 nm long-pass filter prevents resist exposure
during measurement. For those requiring an ultra-violet measurement system,
the optional combined deuterium/halogen lightsource extends this range to
200 nm. A computer-controlled 5-position filter wheel ensures that the wafer
is only exposed to the correct wavelengths. |
Fiberoptics:
Steel-jacketing is
used to protect the plastic clad silica fibers. Standard fiber length is
10’, with longer lengths available as an option (we have provided fiber
runs as long as 150 feet). For ultra-violet applications, high-transmission
silica clad silica fibers reduce UV transmission losses to a minimum. The
standard fiberoptic is a bifurcated bundle for reflectivity measurement,
with custom fiberoptics (for transmission or special lengths) available
with no increase in delivery times. |
| Optical Probes: Several types of optical probes are available. Standard inventory includes working distances from millimeters to feet and spot sizes from 1 to 20 mm as well as right-angle probes for confined working spaces. Probe bodies are made from anodized aluminum, Delrin and stainless steel. Standard optics are designed for the visible, and UV-grade fused silica lenses are used to maximize UV transmission when the optional deuterium/halogen illuminator is ordered. Custom solutions (special space requirements and working distances, spot sizes or vacuum fiberoptic feedthroughs) are available to meet your needs. | Probe
Mounts:
The Inspector 3000 can be used as a bench top instrument or integrated with
the process tool for the in-line and in-situ measurements. The bench top
instrument uses the motorized X-Theta stage for the precise mapping of the
film thickness on the wafer. These measurements are facilitated with the
intuitive Windows operating software. The in-situ and in-line installations
use our compact light probes and fiber optics connecting to the Inspector
hardware outside of the process chamber. The system can be set-up in the
embedded mode with the master controller communicating with the Inspector
through the Serial port. Measurement recipes are kept in the database and
can be easily selected either manually or through the serial communication.
|
| Advanced Develop Rate Applications Module | |
| The
first in a series of data analysis packages designed to increase the process
characterization and development capabilities of the inspection system,
the DevPak Applications Module uses the thickness vs time data as
it’s input. The module makes lithography process development quick and easy.
Dose to clear (EO), contrast and the rate parameter as well as rate parameters
R1, R2, & R3 are calculated in the DevPak. Multiple file formats are supported,
including .csv and ProLith, so the data can be easily exchanged with modeling,
analysis and presentation software. |
| Specifications | ||
IlluminationWavelength |
Halogen Only 380-950 nm |
D2/Halogen
200-950 nm |
SpectrometerRange |
200-1000 nm |
|
Thickness MeasurementNumber of Layers |
|
|
|
Home | Products | About | Contacts | Associated Companies webmaster@sctec.com |